传感器 | 测量范围 | 准确度 | 响应时间 | 分辨率 |
氨氮 | 0-200mg/L | 读数之±10%或2mg/L | - | 0.01mg/L |
氯化物 | 0-1000mg/L | 读数之±15%或5mg/L | - | 0.01mg/L |
硝氮 | 0-200mg/L | 读数之±10%或2mg/L | - | 0.01mg/L |
透气式水位 | 0-10m | ±0.03FS(±0.003m) | - | 0.01mmHg |
气压计 | 375-825mmHg | ±1.5mmHg | - | 0.1mmHg |
蓝绿藻藻蓝蛋白或藻红蛋白 | 0-100ug/L 0-100RFU |
线性:R>0.999,与罗丹明WT连续稀释相关(0-100ug/mL),检出限:0.04ug/LPC | T63<2s | 0.01ug/L; 0.01RFU |
叶绿素 | 0-400ug/L Chl 0-100RFU |
线性:R>0.999,与罗丹明WT连续稀释相关(0-400ug/mL),检出限:0.09ug/LChl | T63<2s | 0.01ug/L Chl; 0.01RFU |
电导率 | 0-200mS/cm | 0-100:读数之±0.5%或0.001mS/cm; 100-200:读数之±1% |
T63<2s | 0.0001-0.01mS/cm (取决于测量范围) |
深度(非透气式) | 0-10m | ±0.04%FS(±0.004m) | T63<2s | 0.001m(取决于测量范围) |
0-100m | ±0.04%FS(±0.04m) | |||
0-250m | ±0.04%FS(±0.10m) | |||
溶解氧-光学 | 0-500%空气饱和度 | 0-200%:读数之±1%或1%; 200-500%:读数之±5% |
T63<5s | 0.1%空气饱和度 |
0-50mg/L | 0-20mg/L:±0.1mg/L或读数之±1%; 20-50mg/L:读数之±5% |
0.01mg/L | ||
fDOM | 0-300ppb(QSE) | 线性:R>0.999,300ppb QS(硫酸奎宁) 溶液连续稀释,检出限:0.07ppb QSE |
T63<2s | 0.01ppb QSE |
ORP | -999-999mV | ±20mV,氧化还原标准溶液 | T63<5s | 0.1mV |
pH | 0-14 | ±0.1,校准温度±10℃范围内 ±0.2,全温度量程 |
T63<3s | 0.01 |
温度 | -5—50℃ | -5—35℃:±0.01℃ 35—50℃:±0.05℃ |
T63<1s | 0.001℃ |
浊度 | 0-4000FNU | 0-999FNU:0.3FNU或读数之±2%; 1000-4000FNU:读数之±5% |
T63<2s | 0-999FNU:0.01FNU 1000-4000FNU:0.1FNU |
以下参数为计算得出参数 | ||||
盐度 | 0-70ppt | 读数之±1.0%或0.1ppt | T63<2s | 0.01ppt |
比电导率 | 0-200mS/cm | 读数之±0.5%或0.001mS/cm | T63<2s | 0.001,0.01,0.1mS/cm (自动选择) |
总溶解固体(TDS) | 0.30-1.00 | - | - | 可变 |
总悬浮固体(TSS) | 0-1500mg/L | - | T63<2s | 可变 |
品牌: | 美国YSI |
型号: | YSI EXO |
加工定制: | 否 |